Purpose:
To evaluate the safety and feasibility of magnetic resonance (MR) imaging at 0.5 T in patients with implanted cardiac pacemakers.
Materials And Methods:
Twenty-one models of pacemakers and 44 pacemaker electrodes were exposed to in vitro MR imaging with continuous registration of pacemaker output and temperature at the lead tip. Prior to MR imaging examination, pacemakers were programmed to an asynchronous mode (A00, V00, or D00). Pacemakers were examined before and after MR imaging. Forty-four patients with implanted pacemakers underwent 51 MR imaging examinations under cardiologic surveillance, continuous electrocardiography, pulse oximetry, and capnographic monitoring.
Results:
MR imaging was safely performed in all patients. None of the pacemakers displayed a pacing dysfunction at MR imaging. No changes occurred in the programmed parameters in any device tested in vivo or in vitro. Maximum increases in the temperature at the lead tips were 8.90 degrees C at a specific absorption rate (SAR) of 0.6 W/kg and 23.50 degrees C under a worst-case radio-frequency (RF) heating condition with an SAR of 1.3 W/kg.
Conclusion:
MR imaging at 0.5 T can be safely performed in patients with implanted pacemakers in carefully selected clinical circumstances when appropriate strategies (programming to an asynchronous mode, adequate monitoring techniques, limited RF exposure) are used.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT