Related Experiment Videos
ESR/Alanine gamma-dosimetry in the 10-30 Gy range
C Fainstein1, E Winkler, M Saravi
1Centro Atómico Bariloche, Rio Negro, Argentina. cfainstein@cab.cnea.gov.ar
Abstract:
We report Alanine Dosimeter preparation, procedures for using the ESR/Dosimetry method, and the resulting calibration curve for gamma-irradiation in the range from 10-30 Gy. We use calibration curve to measure the irradiation dose in gamma-irradiation of human blood, as required in Blood Transfusion Therapy. The ESR/Alanine results are compared against those obtained using the thermoluminescent dosimetry (TLD) method.