Electron arc therapy treatment planning verification with alanine/EPR dosimetry
C De Angelis1, A Mattacchioni, S Onori
1Physics Laboratory, Istituto Superiore di Sanità, Rome, Italy. deangelis@axiss.iss.infn.it
Abstract:
Alanine/EPR dosimetry was applied to measure dose distributions during electron arc therapy treatments generated by electron beams moving isocentrically over a stated arc. Alanine-polyethylene pellets, prepared at ISS according to the NIST recipe, inserted in a homogeneous cylindrical and anthropomorphic phantoms, were used. Preliminary, alanine response to static electron beams in the (6-20) MeV nominal energy range was studied. Then alanine dosimetry was applied to determine the dose versus the gantry angle at a reference point in a cylindrical homogeneous phantom. Finally, arc therapy treatment planning verification was performed in anthropomorphic phantom.