Related Experiment Video
Updated: Jul 28, 2026

Robotized Testing of Camera Positions to Determine Ideal Configuration for Stereo 3D Visualization of Open-Heart Surgery
Published on: August 12, 2021
A localization algorithm and error analysis for stereo x-ray image guidance
1Department of Radiology, Johns Hopkins University, Baltimore, Maryland 21205, USA.
Abstract:
Stereo x-ray radiography attracts increasing attention in major clinical applications. The purpose of this paper is to analyze the 3D localization error for breast biopsy procedures and provide guidelines for improving its accuracy. Our prototype is a CCD based digital stereo x-ray imaging system. The mathematical model consists of two x-ray sources and one stationary detector plane. A closed form least-squares solution is derived for 3D localization of feature points, particularly a biopsy needle tip, from a pair of 2D digital radiographs. Based on the least-squares formula and its first order approximation, the 3D localization error is analyzed in terms of object location, measurement error, separation between the two x-ray sources, and distance from the source to the detector. The stereo imaging and error estimation formulas are numerically simulated and experimentally validated. The data are in agreement with theoretical prediction. These results can be used for the purpose of system design and protocol optimization.
More Related Videos
05:49Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
10:25Technical Approach for Infrared Tracking for Soft Tissue Navigation with a Holographic Head-Mounted Display and Preclinical Validation
Published on: September 2, 2025