The reproducibility of polyacrylamide gel dosimetry applied to stereotactic conformal radiotherapy
V P Cosgrove1, P S Murphy, M McJury
1Joint Department of Physics, The Royal Marsden NHS Trust, Sutton, Surrey, UK. vivian@icr.ac.uk
Abstract:
The reproducibility of polyacrylamide gel (PAG) dosimetry has been evaluated when used to verify two radiotherapy treatment plans of increasing complexity. The plans investigated were a three-field coplanar arrangement, using the linac jaws for field shaping, and a four-field, conformal, non-coplanar plan using precision-cast lead alloy shielding blocks. Each treatment was performed three times using phantoms and calibration gels manufactured in-house. Two phantoms were specially designed for this work to aid accurate positioning of the gels for irradiation and imaging. All gels were imaged post-irradiation using a Siemens Vision 1.5T MR scanner. T2 relaxation images were calibrated to absorbed dose distributions using a number of smaller calibration vessels to produce distribution maps of relative dose. The relative dose distributions were found to be reproducible, with the standard deviation on the mean areas enclosed by the > or = 50% isodose lines measured in three orthogonal planes being 6.4% and 4.1% for the coplanar and non-coplanar plans respectively. The measured distributions were also consistent with those planned, with isodose lines generally agreeing to within a few millimetres. However, the measured absolute doses were on average 23.5% higher than those planned. Although the polyacrylamide gel dosimetry technique has some limitations, particularly when calibrating distributions to absolute dose, the ability to resolve sharp dose gradients in three dimensions with millimetre precision is invaluable when verifying complex conformal treatment plans, where avoidance of proximal, critical structures is a treatment criterion.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014