Differences between blood flow as indicated by the hemodialysis blood roller pump and blood flow measured by an
J L Teruel1, M Fernández Lucas, R Marcén
1Department of Nephrology, Ramón y Cajal Hospital, Madrid, Spain.
Background/Aim:
The ultrasonic transit time is currently the best method for measuring the blood flow rate in the extracorporeal hemodialysis circuit. The purpose of this study was to analyze the differences between blood flow as indicated by the hemodialysis blood roller pump (prescribed blood flow) and by an ultrasonic flowmeter (delivered blood flow).
Methods:
The ultrasonic blood flow was measured in 20 patients on chronic hemodialysis who were dialyzed through an arteriovenous fistula. During each dialysis session the ultrasonic blood flow was measured at three different blood roller pump flow rates (300, 350, and 400 ml/min). In order to analyze the influence of inflow and outflow pressures on blood flow, this study was conducted during nine consecutive dialysis sessions during which needles of different sizes were used.
Results:
The ultrasonic flow was always lower than indicated by the blood roller pump: 265+/-12, 304+/-15, and 341+/- 19 ml/min for blood roller pump flow rates of 300, 350, and 400 ml/min, respectively (variability: -11.6, -13.1, and -14.8%, respectively). An univariate regression analysis showed that the reduction in flow recorded ultrasonically correlated with both venous blood line pressure (r = -0.2679, p<0.001) and negative arterial blood line pressure (r = 0. 6773, p<0.001). By multivariate analysis, only the arterial blood line pressure has a predictive value. When all measurements were grouped by arterial blood line pressure ranges, the variability between ultrasonic blood flow and blood roller pump flow was found to be similar in those groups with the same arterial blood line pressure, regardless of the blood roller pump flow rate.
Conclusions:
The blood flow indicated by the dialysis blood roller pump is always greater than the delivered blood flow, and this difference is in turn conditioned by the negative pressure induced by the blood roller pump in the arterial blood line.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Hemodialysis I: Introduction