[Where should the implantable ECG event recorder be placed?].
E Himmrich1, C Zellerhoff, D Nebeling
1II. Medizinische Klinik Johannes Gutenberg-Universität, Mainz.
Summary
The left parasternal area is the optimal site for implanting an electrocardiogram event recorder (ECG-ER) for superior signal quality. Vertical placement with electrodes toward the skin ensures the best ECG readings.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- Electrocardiogram event recorders (ECG-ER) are crucial for long-term cardiac monitoring.
- Optimal lead placement is essential for high-quality ECG signal acquisition.
- Previous studies have not definitively established the best anterior thoracic site for ECG-ER implantation.
Purpose of the Study:
- To compare bipolar electrocardiogram (ECG) quality at various anterior thoracic sites for ECG-ER implantation.
- To determine the preferable implantation site and orientation for ECG-ERs.
- To evaluate the impact of electrode positioning on ECG signal integrity.
Main Methods:
- Bipolar ECGs were recorded in 70 patients at multiple anterior thoracic sites (subclavicular, parasternal, axilla, apex, subxiphoid) with a 4 cm lead distance.
- ECG quality was assessed based on QRS amplitude, P-wave visibility, pacemaker spike detection, QT interval measurement, and bundle-branch block identification.
- Comparisons included standard Lead II, vertical vs. horizontal electrode positions, and electrode orientation (skin vs. muscle).
Main Results:
- The left parasternal site yielded the best ECG quality and agreement with standard Lead II in 67% of patients.
- Significantly lower quality was observed at right parasternal, left subclavicular, apical, and subxiphoid sites (p < 0.001).
- Vertical electrode positioning resulted in higher QRS amplitude and more visible P-waves compared to horizontal positioning. Electrodes placed towards the muscle caused significant noise.
Conclusions:
- The left parasternal region is the recommended implantation site for ECG-ERs.
- Vertical electrode placement with electrodes oriented towards the skin is advised for optimal signal acquisition.
- These findings can improve the diagnostic yield of ECG event monitoring.