The ESTRO-QUALity assurance network (EQUAL)
I H Ferreira1, A Dutreix, A Bridier
1EQUAL Measuring Laboratory, Service de Physique, Institut Gustave-Roussy, 39 rue Camille Desmoulins, 94805 Villejuif, France. equal@igr.fr
Summary
The European Society for Radiotherapy and Oncology Quality Assurance network (EQUAL) audited 102 centers using mailed thermoluminescent dosimeters (TLD). Results show a small percentage of deviations in radiotherapy dose delivery, highlighting the need for quality assurance.
Area of Science:
- Medical Physics
- Radiation Oncology
- Quality Assurance
Background:
- The European Society for Radiotherapy and Oncology (ESTRO) established the Quality Assurance network (EQUAL) to assess external radiotherapy dose delivery.
- External audits utilize mailed thermoluminescent dosimeters (TLDs) for measurements under reference and non-reference conditions.
Purpose of the Study:
- To evaluate the accuracy of radiation dose delivery in external radiotherapy through a quality assurance program.
- To identify potential deviations in dose output, depth dose, and wedge transmission factors.
Main Methods:
- Thermoluminescent dosimeters (TLDs) made of LiF powder (DTL 937) were read using an automatic TLD reader (PCL 3).
- Participating centers delivered a calculated absorbed dose of 2 Gy to TLDs for up to three photon energies, with 10 on-axis points per beam.
- Key parameters assessed included reference beam output, output variation with collimator settings, depth dose data, and wedge transmission factors.
Main Results:
- In 1998, 102 centers and 235 beams (28 Co-60, 207 X-ray) were audited.
- Approximately 3% of reference beam outputs and percentage depth doses deviated beyond the +/-5% tolerance.
- Deviations of 6% were observed for beam output variation and wedge transmission factors, particularly for large/rectangular fields and wedged beams.
Conclusions:
- The EQUAL program underscores the critical role of quality assurance in radiotherapy, especially for non-reference points on the beam axis.
- Continuous monitoring and adherence to quality standards are essential for ensuring accurate and safe radiation therapy delivery.