[A hybrid technique for the automatic floating table MRA of peripheral arteries using a dedicated phased-array coil
R Janka1, F A Fellner, C Fellner
1Institut für Diagnostische Radiologie, Friedrich-Alexander-Universität Erlangen-Nürnberg.
Purpose:
We introduce a hybrid technique which allows a high resolution MRA of the peripheral arteries with a dedicated phased-array coil using the floating table technique.
Materials And Methods:
Five patients with peripheral arterial occlusive disease were examined within one week with i.a. DSA and MRA using the hybrid technique. MRA examinations were done on a 1.5 T system. At first, pelvic arteries were examined in a single step mode applying the CareBolus technique. Subsequently, thighs and lower legs were examined using the floating table mode. 125 vascular segments were evaluated.
Results:
The hybrid technique proved to be robust and could be performed in each case. Mean examination time was about 30 min. For 117 vascular segments no difference was found between i.a. DSA and MRA. Three segments revealed a higher grade of stenosis in DSA than in MRA, five segments were graded higher in MRA than in DSA. Occlusions were visualized identically in both methods. Venous overlap had no relevant effects on image evaluation.
Conclusions:
This hybrid technique in combination with phased-array coils allows a high resolution MRA of the peripheral arteries with very good image quality. If future studies confirm reduced venous overlap, this method may be an alternative also for users of the floating table MRA with the body resonator.