Recanalization of thrombosed dialysis access with use of a rotating mini-pigtail catheter: follow-up study

T Schmitz-Rode1, J E Wildberger, D Hübner

  • 1Department of Diagnostic Radiology, University of Technology, Uinversitätsklinikum, Aachen, Germany. smiro@rad.rwth-aachen.de

Abstract

Insights

Mechanical thrombectomy using a rotating mini-pigtail catheter effectively declotted dialysis access in all procedures. This safe and simple method offers a low-budget alternative for treating occluded fistulas and grafts.

Area of Science:

  • Vascular Surgery
  • Interventional Nephrology
  • Medical Devices

Background:

  • Dialysis access occlusion is a common complication.
  • Mechanical thrombectomy is a key treatment for restoring access patency.

Purpose of the Study:

  • To assess the feasibility, efficacy, and safety of mechanical thrombectomy for occluded dialysis access using a novel rotating mini-pigtail catheter.

Main Methods:

  • The rotating mini-pigtail catheter mechanically fragmented thrombus in 26 procedures (15 native fistulas, 11 PTFE grafts).
  • Thrombus fragmentation was followed by balloon angioplasty for underlying stenoses.
  • The guide wire acted as a fixed rotation axis for the catheter tip.

Main Results:

  • 100% clinical success rate in declotting dialysis access, enabling immediate use.
  • The mini-pigtail catheter was simple and rapid to handle in both native fistulas and PTFE grafts.
  • No complications related to the thrombus fragmentation procedure were observed.

Conclusions:

  • Mechanical thrombectomy with the mini-pigtail catheter is effective and safe for occluded dialysis access.
  • This technique presents a cost-effective and user-friendly alternative to existing thrombectomy devices.