Related Experiment Video
Updated: Jul 30, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Patient and personnel exposure during CT fluoroscopy-guided interventional procedures
R D Nawfel1, P F Judy, S G Silverman
1Department of Radiology, Brigham and Women's Hospital, 75 Francis St, Boston, MA 02115, USA. nawfel@bwh.harvard.edu
Purpose:
To estimate patient dose and personnel exposure from phantom measurements during computed tomographic (CT) fluoroscopy, to use the estimates to provide users with dose information, and to recommend methods to reduce exposure.
Materials And Methods:
Surface dose was estimated on a CT dosimetric phantom by using thermoluminescent dosimetric (TLD) and CT pencil chamber measurements. Scatter exposure was estimated from scattered radiation measured at distances of 10 cm to 1 m from the phantom. Scatter exposures measured with and without placement of a lead drape on the phantom surface adjacent to the scanning plane were compared.
Results:
Phantom surface dose rates ranged from 2.3 to 10. 4 mGy/sec. Scattered exposure rates for a commonly used CT fluoroscopic technique (120 kVp, 50 mA, 10-mm section thickness) were 27 and 1.2 microGy/sec at 10 cm and 1 m, respectively, from the phantom. Lead drapes reduced the scattered exposure by approximately 71% and 14% at distances of 10 and 60 cm from the scanning plane, respectively.
Conclusion:
High exposures to patients and personnel may occur during CT fluoroscopy-guided interventions. Radiation exposure to patients and personnel may be reduced by modifying CT scanning techniques and by limiting fluoroscopic time. In addition, scatter exposure to personnel may be substantially reduced by placing a lead drape adjacent to the scanning plane.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography