Comparison of erbium:yttrium-aluminum-garnet-laser vitrectomy and mechanical vitrectomy: a clinical study
H Petersen1, M Mrochen, T Seiler
1Department of Ophthalmology, University of Dresden, Dresden, Germany.
Objective:
To evaluate the clinical usefulness of the erbium:YAG (yttrium-aluminum-garnet) laser for vitrectomy and to compare it with a conventional mechanical vitrectomy system regarding the intraoperative parameters.
Design:
Prospective, randomized, clinical trial.
Participants:
Thirty eyes of 30 patients underwent vitrectomy, 15 eyes in each group.
Methods:
For mechanical vitrectomy, a commercially available vitrectomy unit was used. The operating parameters, cutting rate (7 Hz = 420 cpm), maximal suction force (300 mmHg), and aspiration flow (20 ml/min), were held constant. A newly developed erbium (Er):YAG laser unit and handpiece was used for laser vitrectomy with predetermined parameters for cutting rate (70 Hz = 4200 cpm), maximal suction force (50 mmHg), and aspiration flow (20 ml/min). Surgery parameters were recorded in real time and the operation was video recorded. The clinical follow-up time was a minimum of 3 months (average, 6.2 months; range, 3-9 months).
Main Outcome Measurements:
The surgery time was comparable in both groups. During Er:YAG laser vitrectomy, the average suction force was significantly reduced (P< 0.001) compared with that during mechanical vitrectomy. The mean-square variation in suction as a measure to quantify the forces acting on intraocular structures during surgery was significantly smaller in the Er:YAG laser vitrectomy group (P << 0.001).
Conclusions:
Erbium:YAG laser vitrectomy may provide a semicontinuous procedure. This technique minimizes periodical intraoperative forces and movements of intraocular structures and may provide, therefore, a safer vitrectomy.