A study of doses to the hands during dispensing of radiopharmaceuticals
S Dhanse1, C J Martin, T E Hilditch
1Department of Clinical Physics and Bio-Engineering, Western Infirmary, North Glasgow University Hospitals NHS Trust, UK.
Abstract:
Radiation doses to both the fingertip and finger base of staff preparing radiopharmaceuticals in a radionuclide dispensary have been monitored for more than 5 years. The records show a fall in dose thought to be due, largely, to the introduction of all-glass (lead) syringe shields. In 1998, the annual radiation dose (mean +/- standard deviation) to the fingertip had fallen to 83+/-11 mSv, which corresponded to 0.18 mSv per 10 GBq 99Tc(m) handled. There was less reduction in dose to the finger base, which, in 1998, was 70% of that to the fingertip. To study the effect of a syringe shield on the distribution of dose across the hands, model hands were constructed and a gamma extremity monitoring system (GEMS) was used to measure dose. When a radiopharmaceutical is dispensed, contributions to the dose from activity in the vial and syringe change during the procedure. As a result, relative doses to different fingers and the ratios of doses to the tip and base of each finger will fluctuate throughout dispensing. In the absence of a syringe shield, the front tip of the index finger received the greatest radiation dose. When a syringe shield was used, doses were lower and the dose distribution was completely different. These findings have important implications in terms of compliance with the new UK Ionizing Radiations Regulations, where the dose limit is applied to 'the dose averaged over any area of 1 cm2 regardless of the area exposed'.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Biological Effects of Radiation
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Dosage Regimens: Designs and Approaches
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Drug Dosing: Infants and Children