Related Experiment Video
Updated: Aug 14, 2026

High-speed Particle Image Velocimetry Near Surfaces
Published on: June 24, 2013
Accuracy in gamma camera measurements of point source velocities
L C Morgan1, M A Pearson, D W Mackey
1Department of Respiratory Medicine, Concord Repatriation General Hospital, NSW, Australia. lucy@rfu.crg.cs.nsw.gov.au
Abstract:
Mucociliary clearance is impaired in many diseases of the respiratory system. We have developed a method for measuring tracheal mucus velocity by the dynamic study of a single point source of radioactivity deposited in the trachea by cricothyroid injection. Preliminary results suggest that patients with airways disease have very low tracheal mucus velocities (<2 mm x min(-1)). The aim of this experiment was to explore the ability of current scintillation detection systems to track a single point as it moves in a dynamic study in small increments and at low velocity (movements of the order of 1 mm). Background noise was estimated to contribute an error in positioning of 0.16 mm (1 standard deviation). Overall errors in velocity were estimated at 0.2 mm x min(-1). This suggests that standard instrumentation in use in most nuclear medicine departments has the capacity to measure accurately velocities as low as 1 mm x min(-1).