Long-term reliability of semiadjustable articulator interchangeability.

R B Price1, W M Mansfield

  • 1Department of Dental Clinical Sciences, Faculty of Dentistry, Dalhousie University, Halifax, Canada. rbprice@is.dal.ca

Journal of Prosthodontics : Official Journal of the American College of Prosthodontists
|July 15, 2000
PubMed
Summary

Most Whip-Mix model #2240 articulators (83%) remained interchangeable after 1-6 years of clinical use, with minimal vertical discrepancies. This suggests the model #2240 articulator maintains interchangeability over time.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Effects of Light Curing on the Mechanical Properties and Shade Stability of a Dual-Cure Bulk-Fill Resin Composite.

Operative dentistry·2026
Same author

Comparison between the dual-exponential and autocatalytic models to examine rapid photopolymerization kinetics of dental resins.

Dental materials : official publication of the Academy of Dental Materials·2025
Same author

AI-driven dental information.

British dental journal·2025
Same author

Correction to: 3D printing in dentistry: balancing innovation and safety.

British dental journal·2024
Same author

3D printing in dentistry: balancing innovation and safety.

British dental journal·2024
Same author

An Assessment of 1067 Light Curing Units Measured Twice Between 2015 and 2021.

The European journal of prosthodontics and restorative dentistry·2024

Area of Science:

  • Dental Technology
  • Prosthodontics
  • Materials Science

Background:

  • Articulator interchangeability is crucial for accurate transfer of mounted casts in dental procedures.
  • Whip-Mix model #2240 articulators are widely used in clinical settings.

Purpose of the Study:

  • To evaluate the interchangeability of Whip-Mix model #2240 articulators after 1 to 6 years of clinical use.
  • To determine if these articulators maintain accuracy and reliability over time.

Main Methods:

  • Evaluated 46 Whip-Mix model #2240 articulators used clinically for 1-6 years.
  • Assessed horizontal interchangeability using the Whip-Mix #2245 check system (94-micron tolerance).
  • Measured vertical discrepancies using 26-micron shimstock at four locations.

Related Experiment Videos

Main Results:

  • 83% (38 of 46) of articulators passed the interchangeability test (94-micron horizontal tolerance).
  • Most articulators exhibited vertical discrepancies of 26 microns or less.
  • Only 6.5% (3 of 46) showed vertical discrepancies greater than 52 microns.

Conclusions:

  • Whip-Mix model #2240 articulators largely maintain interchangeability within specified tolerances after extended clinical use.
  • The model #2240 articulator demonstrates potential for sustained interchangeability throughout its service life.
  • Clinical use for 1-6 years did not significantly compromise the interchangeability of most Whip-Mix #2240 articulators.