Amplifying and Quantifying HIV-1 RNA in HIV Infected Individuals with Viral Loads Below the Limit of Detection by Standard Clinical Assays
Published on: September 26, 2011
Accuracy of plasma HIV RNA quantification: a multicentre study of variability
S J Skidmore1, M Zuckerman, J V Parry
1Public Health Laboratory, Heartlands Hospital, Birmingham, United Kingdom.
Reproducibility of human immunodeficiency virus (HIV) RNA quantification varied between laboratories and assay kits. Results suggest cautious interpretation of HIV RNA concentration changes for patient treatment decisions.
Area of Science:
- Virology
- Clinical Diagnostics
- Public Health
Background:
- Accurate quantification of plasma human immunodeficiency virus (HIV) RNA concentration is critical for patient care.
- Limited data exists on the reproducibility of HIV RNA quantification within and between clinical laboratories.
Purpose of the Study:
- To assess the reproducibility of plasma HIV RNA concentration measurements across multiple laboratories and assay kits.
- To identify factors influencing variability in HIV RNA quantification.
Main Methods:
- Weekly distribution of identical plasma specimens to 13 laboratories over 13 weeks (January-April 1998).
- Laboratories performed HIV RNA quantification using their routine methods and reported results.
- Analysis of inter-laboratory and inter-assay kit correlation and reproducibility.
Main Results:
- Generally satisfactory correlation of results between laboratories and assay kits.
- HIV RNA concentrations varied by assay kit, with Roche Monitor and AcuGen higher and Chiron Quantiplex v 2.0 lower than average.
- Chiron Quantiplex v 2.0 demonstrated the highest reproducibility; nine 'below detection limit' results were associated with three specimens, including one with HIV-1 clade G RNA.
Conclusions:
- An unsustained change in HIV RNA of
- Performance assessment and conservative interpretation of isolated HIV RNA changes are necessary.
- Parallel testing of consecutive patient specimens offers the most reliable indication of significant HIV RNA concentration change.