Evaluation of portable radionuclide method for measurement of left ventricular ejection fraction and cardiac output
Related Experiment Video
Updated: Aug 11, 2026

Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
Abstract:
Seventeen patients with coronary artery, valvular, or myopathic heart disease were studied to determine correlations of the cardiac output and ejection fraction when comparing the results obtained with a portable probe technique using 113mIn with those obtained with standard methods (cineangiographic, Fick, and dye dilution). With ejection fractions ranging from o.10 to 0.85, the coefficient of correlation was 0.90 when comparing cineangiographic and radionuclide techniques. Cardiac output determinations by the radionuclide technique also correlated well with standard methods (r equals 0.88). The radionuclide method shows promise as an accurate, safe, and simple method in the evaluation of cardiac function at the bedside.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)