Related Experiment Video
Updated: Oct 15, 2026

The Intra-Aortic Balloon Pump
Published on: February 5, 2021
Complications associated with percutaneous placement of intra-aortic balloon counterpulsation: can unsheathed
A Oguzhan1, H L Kisacik, E Varol
1Department of Cardiology, Erciyes University School of Medicine, Kayseri, Turkey. abdurrahmano@hotmail.com
Insights
Sheathless insertion of intra-aortic balloon pump catheters did not reduce lower limb ischemia complications. Peripheral vascular disease, diabetes, and female gender are key risk factors for this complication.
Area of Science:
- Cardiology
- Vascular Surgery
- Medical Devices
Background:
- Percutaneous placement of intra-aortic balloon pump (IABP) counterpulsation is associated with significant complications.
- Despite advancements in catheter design and techniques, complication rates remain high.
- Lower limb ischemia is a primary concern following IABP insertion.
Purpose of the Study:
- To evaluate the effectiveness of a sheathless insertion technique for intra-aortic balloon pump catheters.
- To compare complication rates, specifically lower limb ischemia, between conventional and sheathless insertion methods.
- To identify predictors of lower limb ischemia in patients undergoing IABP insertion.
Main Methods:
- A retrospective study of 148 patients undergoing intra-aortic balloon pumping.
- Patients were divided into two groups: conventional percutaneous insertion (Group I, n=103) and sheathless insertion (Group II, n=45).
- Complication rates, particularly lower limb ischemia, were analyzed and compared between groups. Predictors of limb ischemia were identified.
Main Results:
- The overall complication rate was 16.6% (25 of 150 patients).
- Lower limb ischemia occurred in 11.5% of patients in Group I and 8.9% in Group II, a statistically insignificant difference.
- Peripheral vascular disease, diabetes mellitus, and female gender were significant predictors of limb ischemia (p=0.01, p=0.02, p=0.03, respectively).
Conclusions:
- Sheathless insertion of intra-aortic balloon pump catheters does not significantly reduce the incidence of lower limb ischemia.
- Risk factors such as peripheral vascular disease, diabetes mellitus, and female gender are crucial in predicting limb ischemia after IABP insertion.
- Further research may be needed to explore alternative strategies for reducing IABP-related complications.
Abstract:
Complications related to percutaneous placement of intra-aortic balloon pump counterpulsation are still high despite major refinements in catheter design and techniques. One hundred and forty-eight patients in whom intra-aortic balloon pumping was inserted were classified into two groups on the basis of the insertion technique. Group I included 103 patients in whom the conventional percutaneous insertion was used. A sheathless insertion technique was used in group II (n = 45). The overall complication rate was 16.6% (25 of 150), in which lower limb ischaemia was the most common complication. The limb ischaemia was noted in 12 patients (11.5%) in group I and 4 patients (8.9%) in group II (statistically not significant). Peripheral vascular disease, diabetes mellitus and female gender were found to be significant predictors of limb ischaemia (p = 0.01, p = 0.02 and p = 0.03, respectively). In conclusion, sheathless insertion of intra-aortic balloon pump catheters does not reduce the incidence of limb ischaemia.
Related Concept Videos
Peripheral Artery Disease V: Postoperative Nursing Management
Peripheral Artery Disease III: Interprofessional Care
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization
Aneurysm III: Interprofessional Care
Cardiac Catheterization III: Left Heart Catheterization