Related Experiment Video
Updated: Aug 12, 2026

10:05
Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
Experience with a higher impedance, fixed helix, steroid-eluting pacing lead. The SWEET PICOTIP Rx Study Group
M C Giudici1, M Lee, S Higgins
1Division of Cardiology, Genesis Medical Center, Davenport, Iowa, USA. Giudici@CVMedPC.com
Pacing and Clinical Electrophysiology : PACE
|July 29, 2000
Summary
This study shows a new steroid-eluting pacing lead offers improved electrical performance. The lead demonstrated lower pacing thresholds and higher impedance, potentially reducing energy needs for pacemakers.
Area of Science:
- Cardiovascular Medicine
- Biomedical Engineering
- Medical Devices
Background:
- Pacemaker leads are crucial for delivering electrical impulses to the heart.
- Steroid-eluting leads aim to reduce inflammation and improve lead performance.
- Optimizing lead impedance and pacing thresholds is essential for efficient pacemaker function.
Purpose of the Study:
- To evaluate the electrical characteristics of a novel steroid-eluting pacing lead.
- To compare the performance of the study lead against historical control data.
- To assess the impact of the lead's design on pacing system energy requirements.
Main Methods:
- A multicenter trial involving the implantation of 132 pacing leads in 61 patients.
- Comparison of electrical parameters (pacing threshold, impedance, sensing) with historical control data.
- Analysis of lead performance at the 3-month follow-up point.
Main Results:
- The study lead exhibited significantly lower pacing thresholds in both the atrium (0.72 vs. 1.29) and ventricle (0.71 vs. 1.52) (P < 0.01).
- Significantly higher impedance values were recorded for the study lead in the atrium (923 vs. 719) and ventricle (980 vs. 664) (P < 0.01).
- Cardiac sensing capabilities were comparable between the study lead and historical controls.
Conclusions:
- The steroid-eluting pacing lead demonstrates superior electrical properties, including lower pacing thresholds and higher impedance.
- These characteristics suggest a potential for reduced energy consumption in pacemaker systems.
- The findings support the clinical utility of this advanced pacing lead technology.

