Abstract:
Thrombus formation and hemolysis have been linked to the dynamic flow characteristics of heart valve prostheses. To enhance our understanding of the flow characteristics past the aortic position of a Jellyfish (JF) valve in the left ventricle, in vitro laser Doppler anemometry (LDA) measurements were carried out under physiological pulsatile flow conditions. The hemodynamic performance of the JF valve was then compared with that of the St. Vincent (SV) valve. The comparison was given in terms of mean systolic pressure drop, back flow energy losses, flow velocity, and shear stresses at various locations downstream of both valves and at cardiac outputs of 3.5 L/min, 4.5 L/min, and 6.5 L/min respectively. The results indicated that both valves created disturbed flow fields with elevated levels of turbulent shear stress as well as higher levels of turbulence in the immediate vicinity of the valve and up to 1 diameter of the pipe (D) downstream of the valve. At a location further downstream, the JF valve showed better flow characteristics than the SV in terms of velocity profiles and turbulent shear stresses. The closure volume of the SV valve was found to be 2.5 times higher than that of the JF valve. Moreover, the total back flow losses and mean systolic pressure drop also were found to be higher in the SV than the JF valve.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation.
Aortic Regurgitation I: Introduction
Aortic Regurgitation II: Clinical Features and Diagnostic Tests