A comparison of the effect of task-specific with fixed beam alignment in tuned-aperture computed tomography
1Wake Forest University School of Medicine, Department of Dentistry, Medical Center Boulevard, Winston-Salem, NC 27157-1093, USA.
Objectives:
To test the hypothesis that fixed beam alignment, representative of conventional, fixed-projection three-dimensional (3D) imaging systems, yields significantly lower detectability of approximal radiolucencies than does a task-specific strategy that 'opens the contacts'.
Methods:
Eleven pairs of extracted teeth containing systematically varying numbers of induced approximal low-contrast defects were evaluated from 3D displays produced by means of tuned-aperture computed tomography (TACT) under controlled conditions. Each of the resulting tomographic presentations was displayed in two random sequences to seven trained clinicians. The task-specific (test) series was generated from eight different projections varying in vertical disparity but constrained to open the contacts. The fixed-aiming (control) series was generated from eight projections distributed uniformly on a circle.
Results:
Accuracy of defect detection averaged 5%, significantly higher with task-specific aiming than with fixed alignment projection geometry (P = 0.02). All observers unequivocally preferred the displays produced from projections with open contacts; however, they demonstrated considerable variation in their approach to the task, including the time required to interpret the displays.
Conclusions:
The improvement in observer accuracy obtained from radiographs displayed in three dimensions results from an imaging strategy that selectively increases the signal-to-noise ratio in regions of interest.