Changes in chest electrode impedance.

F Lateef1, S H Lim, V Anantharaman

  • 1Department of Emergency Medicine, Singapore General Hospital and the Nanyang Technological University, Mechanical and Production Engineering. f_lateef@hotmail.com

The American Journal of Emergency Medicine
|August 5, 2000
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Asymptomatic lymphatic filariasis in an elderly patient from Bako, Sarawak: A case report and public health implications.

Tropical biomedicine·2026
Same author

Real-world characteristics and outcomes of patients with BRAF<sup>V600E</sup>-mutant metastatic colorectal cancer in Australia: the COALA project.

ESMO real world data and digital oncology·2026
Same author

Advancing breast cancer research through real-world data from the Singapore Joint Breast Cancer Registry.

ESMO real world data and digital oncology·2026
Same author

Long-Range Transverse-Momentum Correlations and Radial Flow in Pb-Pb Collisions at the LHC.

Physical review letters·2026
Same author

Search for Quasiparticle Scattering in the Quark-Gluon Plasma with Jet Splittings in pp and Pb-Pb Collisions at sqrt[s_{NN}]=5.02  TeV.

Physical review letters·2025
Same author

Knowledge and Skills of Leprosy Among Healthcare Workers and Its Associated Factors in Primary Healthcare.

Inquiry : a journal of medical care organization, provision and financing·2025
Same journal

Rocuronium dosing for rapid sequence intubation: A retrospective analysis in ED and ICU settings.

The American journal of emergency medicine·2026
Same journal

Clinical features of adults with undiagnosed acute leukemia in the emergency department: A descriptive study.

The American journal of emergency medicine·2026
Same journal

Occult intraperitoneal gynecological hemorrhage: An under-evaluated source of hemoperitoneum after blunt trauma.

The American journal of emergency medicine·2026
Same journal

Racket and paddle sports-related injuries treated in United States emergency departments, 2000-2023.

The American journal of emergency medicine·2026
Same journal

Prehospital epinephrine as a bridge to survival in traumatic cardiac arrest: A nationwide propensity score-matched analysis.

The American journal of emergency medicine·2026
Same journal

Crowdsourcing technologies for out-of-hospital cardiac arrest response: A scoping review.

The American journal of emergency medicine·2026
See all related articles

Electrode skin impedance (ESI) decreases over time and is lower at lower frequencies. Optimizing ESI with lower frequencies may enhance ventricular fibrillation (VF) amplitude for automated external defibrillators.

Area of Science:

  • Cardiology
  • Biomedical Engineering
  • Medical Devices

Background:

  • The Prehospital Defibrillation Program in Singapore has observed lower ventricular fibrillation (VF) amplitudes than typical.
  • Electrode skin impedance (ESI) is a critical factor influencing defibrillator performance.
  • Understanding ESI variations is essential for optimizing automated external defibrillator (AED) efficacy.

Purpose of the Study:

  • To measure ESI in patients at 5 Hz and 2 kHz frequencies.
  • To assess ESI changes over time after electrode application.
  • To evaluate ESI differences between electrode positions and correlate with patient factors.

Main Methods:

  • Prospective study measuring ESI in 72 adult patients (≥25 years, not critically ill).

Related Experiment Videos

  • ESI was measured using a modified Heart-Save 911 defibrillator at 10 seconds, 1 minute, and 2 minutes post-electrode application.
  • Measurements were taken at two distinct electrode placement positions using 5 Hz and 2 kHz frequencies.
  • Main Results:

    • ESI significantly decreased over time after electrode application at both frequencies and positions (P < .05).
    • ESI was significantly lower at 5 Hz compared to 2 kHz.
    • No significant correlation was found between ESI and patient's body weight or sex.
    • No statistically significant difference in ESI was observed between the two electrode placement positions.

    Conclusions:

    • Lowering measurement frequency can reduce ESI, potentially enhancing VF amplitude recorded by AEDs.
    • ESI is dynamic and decreases post-application, suggesting a need for standardized measurement timing.
    • Optimizing electrode-skin interface through frequency selection may improve prehospital defibrillation outcomes.