Chest radiography with a large-area detector based on cesium-iodide/amorphous-silicon technology: image quality and dose requirement in comparison with an asymmetric screen-film system
Related Experiment Video
Updated: Oct 15, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
1Department of Diagnostic Radiology, University Hospital Regensburg, Germany.
Abstract:
The purpose of this study was to evaluate a large-area, flat-panel X-ray detector, which uses cesium-iodide (CsI) and amorphous silicon (a-Si). Conventional images were compared with digital images acquired with equal dose (2.5 microGy) and with 50% dose reduction. Fifteen consecutive patients were studied prospectively using an asymmetric screen-film system (detector dose, 2.5 microGy). Digital images were taken from the same patients in a posteroanterior view with detector doses of 2.5 and 1.25 microGy, respectively. The CsI/a-Si active-matrix imager had a panel-size of 43 x 43 cm, a matrix of 3 x 3k, and a pixel-pitch of 143 microm. Hard copies were presented in a random order to eight independent observers, who rated image quality according to six subjective quality criteria. Statistical significance of differences was evaluated with Student's t test for paired samples (confidence level, 95%). Digital radiographs with 2.5 and 1.25 microGy were superior to conventional images regarding all quality criteria. Statistically significant differences were observed for five of six criteria at a detector dose of 2.5 microGy and for only one quality feature at 1.25 microGy. Flat-panel digital imagers based on CsI/a-Si technique have the potential to replace conventional systems and might allow a reduction of radiation dose by 50% without loss of image quality.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Determination of Crystal Structures