Related Experiment Video
Updated: Aug 10, 2026

A Novel Technique for Raman Analysis of Highly Radioactive Samples Using Any Standard Micro-Raman Spectrometer
Published on: April 12, 2017
Comparison and improvement of the determinations of actinide low activities using several alpha liquid scintillation
1Groupe de Radiochimie, Institut de Physique Nucleaire, Universite Paris-Sud, Orsay, France. dacheux@ipno.in2p3.fr
Abstract:
We applied three procedures using two a liquid scintillation spectrometers (PERALS and TRI-CARB) and two scintillation cocktails (Alphaex and Ultima Gold LLT) for the determination of alpha-emitter low activities. For each procedure, the limit of detection, the resolution, the separation factor, and the Fischer coefficient were determined in order to perform 232U-234U-238U isotopic measurements. The deconvolution usually performed is clean when the PERALS spectrometer is used. This is not possible for the TRI-CARB spectrometer using the Ultima Gold LLT scintillation cocktail. This problem was solved by combining the advantages of both techniques using the Alphaex scintillation cocktail in the TRI-CARB spectrometer. Under these conditions, the limit of detection was improved, the resolution decreased from 500-800 to 420-590 keV, and the separation factor increased from 0.9 to 1.1-1.2. This third procedure was applied with success for 232U-234U-238U isotopic experiments.
Related Concept Videos
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Lab
Atomic Fluorescence Spectroscopy