CO assessment by suprasternal Doppler in critically ill patients: comparison with thermodilution
Related Experiment Video
Updated: Oct 13, 2026

Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
J Y Lefrant1, M Benbabaali, J Ripart
1Fédération d'Anesthésie Douleur Urgences Réanimation, Centre Hospitalier Universitaire de Nîmes, France.
Objective:
Comparison of suprasternal Doppler (SST) and thermodilution (TD) for the measurement of cardiac output (CO) in critically ill patients.
Design:
Prospective study.
Setting:
Intensive care unit of a university hospital.
Patients And Participants:
65 consecutive critically ill patients requiring a pulmonary artery catheter.
Interventions:
Paired CO measurements were made simultaneously using SST and TD by two independent operators. The time to obtain a CO value by SST was measured. Correlation coefficients and the linear regression equation were determined. A Bland and Altman diagram was plotted. A Bland and Altman diagram was also plotted for the level of cardiac index (CI) values (low: CI < 2.5 l min(-1) m(-2); normal: 2.5 < or = CI < or = 4.5 l min(-1) m(-2); high: CI > 4.5 l min(-1) m(-2)).
Measurements And Results:
In seven patients SST failed to measure CO. In the remaining 58 patients 314 paired CO measurements were performed. The mean time to measure CO by SST was 73 +/- 45 s. The equation of linear regression was: SST(CO) = 0.84 TD(CO) + 1.39. The correlation coefficient was 0.84. The bias between SST and TD was -0.2 +/- 1.4 l min(-1). Biases were -0.23 +/- 0.50, -0.20 +/- 0.68, and 0.25 +/- 0.92 l min(-1) m(-2) for low, normal, and high levels of CI, respectively.
Conclusion:
SST does not accurately measure CO but allows a rapid assessment of CI level in critically ill patients.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Acute Coronary Syndrome III: Diagnostic Studies
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.