The strength-duration curve and its importance in pacing efficiency: a study of 325 pacing leads in 229 patients
1Department of Cardiology, Wansbeck General Hospital, Ashington, Northumberland, United Kingdom. coates_steve@hotmail.com
Abstract:
Pacemaker battery life is dependent on programmable parameters, principally pulse amplitude and pulse duration. High factory default settings cause excessive current drain. The strength-duration curve relates pacing threshold to pulse duration. The most energy efficient pacing occurs at chronaxie, a value of pulse duration derived from the curve. Strength-duration curves were calculated for 325 acutely implanted pacing leads. Chronaxie and rheobase were compared for atrial and ventricular leads. Chronaxie was compared with actual programmed pulse duration. There were 101 atrial and 224 ventricular leads, all passive fixation. The curve fit was good, (mean error +/- SD) 0.024 +/- 0.06 V for atrial curves and 0.008 +/- 0.034 V for ventricular curves. Mean (+/- SD) atrial and ventricular chronaxies were 0.24 +/- 0.07 ms and 0.25 +/- 0.07 ms, respectively. A "Z" value of 1.4 indicated that chronaxies might have been from the same population. Mean (+/- SD) atrial and ventricular rheobases were 0.51 +/- 0.2 V and 0.35 +/- 0.13 V, respectively. A "Z" value of 7.1 (P < 0.001) suggested atrial and ventricular rheobases were from differing populations. All patients had factory default pulse durations of 0.45 ms or 0.5 ms, exceeding acute chronaxie by a factor of two, thus, demonstrating suboptimal pacing. We conclude that understanding the strength-duration curve is critical. Sensible programming of other pacing functions optimizes longevity. Battery drain is reduced by programming pulse duration to chronaxie with a doubling of voltage threshold at this point to achieve a safety margin. Further study of chronaxie drift with time is required.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electrophysiology of Normal Cardiac Rhythm
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Dysrhythmias V: Evaluating Dysrhythmias