Related Experiment Video
Updated: Aug 15, 2026

A Cost-effective and Reliable Method to Predict Mechanical Stress in Single-use and Standard Pumps
Published on: August 5, 2015
Physical model-based indirect measurements of blood pressure and flow using a centrifugal pump
T Kitamura1, Y Matsushima, T Tokuyama
1Department of Mechanical System Engineering, College of Computer Science and Systems Engineering, Kyushu Institute of Technology, Fukuoka, Japan. kita@mse.kyutech.ac.jp
Abstract:
This article describes a technique offering indirect measurements of pump pressure differential and flow with certain accuracy independent of changes in blood viscosity. This technique is based on noninvasive measurements of the motor current and rotation speed using the physical model equations of the centrifugal pump system. Blood viscosity included in the coefficients of the dynamic equations is first estimated, and then substitution of the estimated viscosity into the steady equations of the model provides pump flow and pressure differential. In vitro tests using a Capiox pump showed a sufficient linear correlation between actual values and their estimates for pressure differential and pump flow. An in vivo test using a 45 kg sheep showed that the proposed algorithm needs robustness for the convergence of estimates of viscosity. An overall evaluation, however, of the developed algorithm/model showed indications of success in terms of efficient computation and modeling.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Blood Pressure
Measurement of Blood Pressure
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...