Related Experiment Video
Updated: Aug 9, 2026

Use of Two Intracorporeal Ventricular Assist Devices As a Total Artificial Heart
Published on: May 11, 2018
Continuously maintaining positive flow avoids endocardial suction of a rotary blood pump with left ventricular bypass
T Akimoto1, K Yamazaki, P Litwak
1Department of Surgery, University of Pittsburgh, Pittsburgh, Pennsylvania 15219, USA.
Abstract:
This study showed the usefulness of maintaining positive pump flow to avoid endocardial suction and as an assist bypass. Three calves were implanted with centrifugal pumps. Hemodynamics and pump parameters were measured at varying pump speeds (from 1,100 to 2,300 rpm). In each test pump, speed was adjusted to create 3 hemodynamic states: both positive and negative flow (PNF), positive and zero flow (PZF), and continuously positive flow (CPF). The pump flow volume was determined during systole (Vs) and diastole (Vd). Vs in PNF was 29.6 ml and was not significantly different from Vs in PZF (p > 0.15). Vd in PNF was significantly different from Vd in PZF (p < 0.05). All bypass rates of PNF were over 30% of pulmonary flow. All PZF bypass rates were between the PNF rate and the CPF rate. These data showed that PZF satisfied the minimum requirement of assist flow and was under 100% bypass. Thus, PZF may avoid endocardial suction.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Cardiac Cycle
During the cardiac cycle, blood flow through the heart is regulated entirely by changing pressure gradients. This sequence of events begins with the heart in a state of total relaxation, known as mid-to-late diastole, during which blood passively flows from...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Mitral Stenosis I: Introduction
Heart Failure VI: Adjunct Therapies
Cardiomyopathy V: Interprofessional Care