Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
Biocompatibility of heparin-coated extracorporeal bypass circuits: new heparin bonded bioline system
E Tayama1, N Hayashida, K Akasu
1Department of Surgery, Kurume University School of Medicine, Kurume, Japan. eiki@med.kurume-u.ac.jp
Abstract:
Biocompatibility of a new type of heparin-coated cardiopulmonary bypass equipment, the Bioline, was evaluated in coronary artery bypass surgery cases. The heparin-coated (H) group (n = 15; Quadrox Bioline oxygenator/reservior and Carmeda BioMedicus BP-80 centrifugal pump) was compared with the nonheparin-coated (N) group (n = 12; uncoated, otherwise similar oxygenator, centrifugal pump, tubing, and filter set). Both groups used full systemic heparinization. The peak values of neutrophil elastase, C3a, IL-6, and IL-8 at 2 h after cardiopulmonary bypass (CPB), and C3a levels at the end of CPB and at 2 h after CPB were significantly reduced in the H group compared with those of the N group. However, no statistically significant intergroup differences were observed in thrombin-antithrombin complex, D-dimer, beta-thromboglobulin, or platelet factor-4. No significant differences were observed in hemostasis time, postoperative 12 h blood loss, required amount of blood transfusion, or intubation time. In conclusion, the Bioline demonstrated partially improved biocompatibility, in terms of leukocyte and complement activation, and proinflammatory cytokine production. However, it did not improve platelet activation, coagulation, or fibrinolysis cascade under full systemic heparinization. As a result, the clinical beneficial impact seemed to be the minimum.
More Related Videos
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
10:15The Use of the Ex Vivo Chandler Loop Apparatus to Assess the Biocompatibility of Modified Polymeric Blood Conduits
Published on: August 20, 2014
Related Concept Videos
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Extrinsic and Intrinsic Pathways of Hemostasis
The Extrinsic Pathway
The extrinsic pathway of coagulation is typically initiated by tissue damage that exposes blood to tissue factor (TF), a protein released by the damaged tissue cells outside the blood vessels—this interaction with TF triggers biochemical reactions involving specific clotting factors. The key player here is Factor VII, which forms a...