Factitious meningitis. Diagnostic error due to nonviable bacteria in commercial lumbar puncture trays
Abstract:
A cluster of five false-positive cerebrospinal fluid (CSF) Gram stains led to an investigation of possible causes of specimen or smear contamination. Specimen tubes supplied in commercial lumbar puncture trays from the lot being used in the involved hospital were shown to contain nonviable bacteria, When filled with a test solution and processed in a manner similar to that used for processing CSF specimens, 10 of 12 tubes evaluated yielded Gram stains containing either Gram-negative rods, diplococci, or coccobacilli. Before this problem was recognized, the patient from whom the first false-positive smear was obtained was treated for bacterial meningitis. It is important to realize that microbial contamination of commercial CSF specimen tubes can result in findings simulating those of early bacterial meningitis.
Insights
Contaminated cerebrospinal fluid (CSF) specimen tubes can cause false-positive Gram stains, mimicking bacterial meningitis. This contamination necessitates careful evaluation of diagnostic materials to ensure accurate patient care.
Area of Science:
- Clinical Microbiology
- Diagnostic Pathology
- Infectious Diseases
Background:
- A series of five false-positive cerebrospinal fluid (CSF) Gram stains prompted an investigation into potential sources of contamination.
- The clinical implications of false-positive results in diagnosing serious infections like bacterial meningitis are significant.
Purpose of the Study:
- To identify the cause of recurrent false-positive Gram stain results in CSF specimens.
- To determine if commercial lumbar puncture trays were the source of bacterial contamination.
Main Methods:
- Commercial lumbar puncture trays from the implicated lot were tested.
- Specimen tubes were filled with a test solution and processed similarly to CSF samples.
- Gram stains were performed on the processed solutions to detect microbial presence.
Main Results:
- Nonviable bacteria were detected in specimen tubes from the commercial lumbar puncture trays.
- Ten out of twelve evaluated tubes, when processed, yielded Gram stains with Gram-negative rods, diplococci, or coccobacilli.
- A patient initially treated for bacterial meningitis due to a false-positive result.
Conclusions:
- Microbial contamination of commercial CSF specimen tubes is a potential source of diagnostic error.
- Findings from contaminated tubes can simulate early bacterial meningitis, leading to misdiagnosis and inappropriate treatment.
- Healthcare facilities should be vigilant regarding the sterility of diagnostic supplies to prevent patient harm.