Objective:
To evaluate the accuracy of the pulse oximeter to detect hypoxemia in patients with sickle cell disease in an ambulatory care setting.
Study Design:
Simultaneous measurements of PaO(2), arterial oxygen saturation by co-oximetry (criterion standard), and pulse oximetry were performed in 21 children with sickle cell disease during 22 outpatient visits. The bias and precision of the pulse oximeter compared with measured arterial oxygen saturation by co-oximetry were determined. The sensitivity, specificity, and positive and negative predictive values of the pulse oximeter to detect hypoxemia (PaO(2) <70 mm Hg) were also calculated.
Results:
The mean difference between pulse oximetry and measured oxygen saturation (bias) was 5.0% and the SD (precision) was 5.3. Twenty-one patients had a PaO(2) greater than 70 mm Hg; 7 of these (33%) were predicted to be hypoxic by pulse oximetry with values less than 93%, for a specificity to detect normoxia of 67%.
Conclusion:
Making treatment decisions based on pulse oximetry data alone in patients with sickle cell disease who are not acutely ill may be inappropriate.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Pulse Oximetry
Purpose