Finite Element Analysis Model for Assessing Expansion Patterns from Surgically Assisted Rapid Palatal Expansion
Published on: October 20, 2023
The biomechanics of rapid maxillary sutural expansion
S Braun1, J A Bottrel, K G Lee
1Clinical Professor of Orthodontics, Vanderbilt University Medical Center, Indianapolis, IN 46240, USA. ortho.braun@juno.com
Abstract:
Micro-displacements (fringe patterns) in the bones of the craniofacial complex as seen through laser holography during midpalatal sutural expansion with the Hyrax appliance are used to define the centers of rotation of the maxillary halves in both the frontal and occlusal views. Biomechanical analyses of the maxillary expansion force system are concomitant with the holographic findings and strongly suggest that the stainless steel wires joining the teeth to any expansion device be of the largest diameter possible. In addition, in the case of the Hyrax expansion device, it is recommended that the manufacturer increase the diameter of the activating screw as well as those of the 2 adjacent wire guides. And, importantly, the use of acrylic as a structural member to join the teeth to a sutural expansion device should be avoided if tipping of the maxillary halves is to be minimized, as the acrylic lacks sufficient rigidity.