Determination of clemastine in human plasma by gas chromatography with nitrogen-phosphorus detection
N N Davydova1, S U Yasuda, R L Woosley
1Department of Pharmacology, School of Medicine, Georgetown University, Washington, DC 20007, USA.
Abstract:
A method for the quantitative determination of clemastine in human plasma has been developed and validated. The assay uses gas chromatography with nitrogen-phosphorus detection and a HP-1 capillary column (25 mx0.22 mm, film thickness 0.33 mm) coated with dimethylpolysiloxane. Clemastine (with orphenadrine as internal standard) was isolated from human plasma using liquid-liquid extraction. A linear relationship was observed between 0.1 and 12.8 ng/ml using the peak area ratio of clemastine to orphenadrine with a correlation coefficient greater than 0.99 (the detection limit for clemastine was 0.06 ng/ml). The intra- and inter-day coefficients of variation were less than 11%. The developed method was used for the analysis of plasma samples from healthy volunteers (n = 19) to examine the pharmacokinetics of the antihistamine clemastine after single and multiple oral doses of clemastine fumarate.
More Related Videos
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II