Related Experiment Video
Updated: Aug 2, 2026

Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Extended use of the k wire and the orthopaedic screw: the fixclip project
1Northampton General Hospital, England, UK. allenbaker@uk-consultants.co.uk
Abstract:
The results of the first 280 fixations with the Fixclip systems are reported. The intended advantages of increased fixation accuracy and versatility have been realised. Two hundred and fifty four fixations have been followed-up to union; there have been three deep infections, two persistent non-unions, and 12 fixation failures requiring revision. A range of sizes has allowed use of the system with screws from the small and basic fragment sets with wires from 1.2 to 3 mm diameter. Its use in the management of paediatric conditions, intra-articular fractures, fractures in osteoporotic bone and as a blade plate substitute are illustrated. Its potential as a biologically and mechanically effective means of bone fixation is discussed.
More Related Videos
05:34A Mini-Invasive Internal Fixation Technique for Studying Immobilization-Induced Knee Flexion Contracture in Rats
Published on: May 20, 2019
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025