Related Experiment Video
Updated: Aug 14, 2026

A System to Create Stable Nanoparticle Aerosols from Nanopowders
Published on: July 26, 2016
Characteristics of an aerosol photometer while automatically controlling chamber dilution-air flow rate
P T O'Shaughnessy1, D R Hemenway
1Department of Occupational and Environmental Health, University of Iowa, 100 Oakdale Campus, #180 IREH, Iowa City, IA 52242-5000, USA. patrick-oshaughnessy@uiowa.edu
Abstract:
Trials were conducted to determine those factors that affect the accuracy of a direct-reading aerosol photometer when automatically controlling airflow rate within an exposure chamber to regulate airborne dust concentrations. Photometer response was affected by a shift in the aerosol size distribution caused by changes in chamber flow rate. In addition to a dilution effect, flow rate also determined the relative amount of aerosol lost to sedimentation within the chamber. Additional calculations were added to a computer control algorithm to compensate for these effects when attempting to automatically regulate flow based on a proportional-integral-derivative (PID) feedback control algorithm. A comparison between PID-controlled trials and those performed with a constant generator output rate and dilution-air flow rate demonstrated that there was no significant decrease in photometer accuracy despite the many changes in flow rate produced when using PID control. Likewise, the PID-controlled trials produced chamber aerosol concentrations within 1% of a desired level.
Related Concept Videos
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Emission Spectroscopy: Instrumentation
Flame Photometry: Overview