HPLC Measurement of the DNA Oxidation Biomarker, 8-oxo-7,8-dihydro-2’-deoxyguanosine, in Cultured Cells and Animal Tissues
Published on: August 1, 2015
Chromatographic determination of 8-oxo-7,8-dihydro-2'-deoxyguanosine in cellular DNA: a validation study
1Université Libre de Bruxelles, Institute of Pharmacy, Laboratory of Bioanalytical Chemistry, Toxicology and Applied Physical Chemistry, Brussels, Belgium. pduez@ulb.ac.be
Abstract:
Although a series of biomarkers are widely used for the estimation of oxidative damage to biomolecules, validations of the analytical methods have seldom been presented. Formal validation, that is the study of the analytical performances of a method, is however recognized as the best safeguard against the generation and publication of data with low reliability. Classical validation parameters were investigated for the determination of an oxidative stress biomarker, 8-oxo-7,8-dihydro-2'-deoxyguanosine (8-oxo-dG) in cellular DNA, by high-performance liquid chromatography coupled to amperometric detection (HPLC-EC); this modified base is increasingly considered as a marker of oxidative damage to DNA, but many questions are still raised on the analytical methods in use. Upon a rigorous statistical evaluation of the quality criteria currently required for assays in biological media, including selectivity, linearity, accuracy, repeatability, sensitivity, limits of detection and quantification, ruggedness and storage at different stop points in the procedure, the HPLC-EC assay method is found mostly reliable. The present validation attempt demonstrates that (i) the HPLC-EC assay of 8-oxo-dG provides consistent data allowing to reliably detect an increase of this biomarker in cellular DNA; (ii) a harsh oxidative stress does not hinder the enzymatic digestion of DNA by nuclease P1; and (iii) the analytical results must be expressed relative to the internal standard dG which significantly improves both repeatability and sensitivity. Whereas the described assay minimizes the artifactual production of the analyte from processing and storage, this cannot be totally ruled out; the true 8-oxo-dG base levels still lack a definitive assay method, which remains a considerable analytical challenge and the object of controversy.
More Related Videos
12:15Quantification of three DNA Lesions by Mass Spectrometry and Assessment of Their Levels in Tissues of Mice Exposed to Ambient Fine Particulate Matter
Published on: May 29, 2019
05:13Quantifying the Level of 8-oxo-dG Using ELISA Assay to Evaluate Oxidative DNA Damage in MCF-7 Cells
Published on: May 24, 2024