Smaller-than-expected sevoflurane concentrations using the Sevotec 5 vaporizer at low fill states and high fresh gas
1Department of Anesthesiology, Doernbecher Children's Hospital, Oregon Health Sciences University, Portland, Oregon 97201, USA. seropian@ohsu.edu
Abstract:
Smaller-than-expected concentrations of sevoflurane were delivered by Sevotec 5 vaporizers (Datex-Ohmeda, Madison, WI) at low-fill states, 8% dialed concentrations, and high fresh gas flows. Clinically, this would lead to prolonged induction times. The risk of complications from prolonged inhaled induction could be increased when using a vaporizer with these characteristics.