Related Experiment Videos
Wall correction factors for calibration of plane-parallel ionization chambers with high-energy photon beams
F Araki1, R Ikeda, Y Shirakawa
1Department of Radiological Technology, Kumamoto University College of Medical Science, Japan.
Abstract:
Most dosimetry protocols recommend that calibration of plane-parallel ionization chambers be performed in an electron beam of sufficiently high energy by comparison with cylindrical chambers. For various plane-parallel chambers, the 1997 IAEA TRS-381 protocol includes an overall perturbation factor pQ for electron beams, a wall correction factor p(wall) for a 60Co beam and the product of two wall corrections k(att)k(m) for 60Co in-air calibration. The recommended values of p(wall) for plane-parallel chambers, however, are limited to certain phantom materials and a 60Co beam, and are not given for other phantom materials and x-ray beams. In this work, the p(wall) values of the commercially available NACP, PTW/Markus and PTW/Roos plane-parallel chambers in a solid water phantom have been determined with 60Co and 4 and 10 MV photon beams. The k(att)k(m) values for the NACP and PTW/Markus chambers have also been obtained. The wall correction factors p(wall) and k(att)k(m) have been determined by intercomparison with a calibrated Farmer chamber. The average value of p(wall) for these plane-parallel chambers was 1.005 +/- 0.1% (1 SD) for 60Co beams and 1.007 +/- 0.2% (1 SD) for both 4 MV and 10 MV photons. The k(att)k(m) values for the NACP and PTW/Markus chambers were about 1.5% lower than other published data.