Evaluation of Left Ventricular Structure and Function using 3D Echocardiography
Published on: October 28, 2020
Evaluation of left ventricular volumes and ejection fraction with a nonfluoroscopic endoventricular three-dimensional
G Van Langenhove1, J N Hamburger, P C Smits
1Thoraxcenter Rotterdam, Dijkzigt Hospital, Erasmus University, The Netherlands.
Background:
Recently, a novel nonfluoroscopic 3-dimensional electromechanical mapping technique was introduced in the clinical arena. Although initial in vitro and in vivo studies suggested the reliability of the system in volumetric and hemodynamic evaluation of the left ventricle, no validation in human beings has been performed.
Methods:
A nonfluoroscopic electromechanical mapping (NOGA, Biosense-Webster) procedure was performed in 44 patients. All patients received a contrast left ventriculogram during the same session. Volumetric (end-diastolic [EDV] and end-systolic volumes [ESV]) and hemodynamic (left ventricular ejection fraction [LVEF] and stroke volume) parameters of both systems were compared.
Results:
Two uncomplicated pericardial effusions occurred with the first-generation mapping catheters. No procedural complications were noted with the new-generation mapping catheters. Significant correlations were found between mapping-derived and ventriculography-based measurements for both ESV (r = 0.67, P <.001) and LVEF (r = 0.78, P <.001). Absolute volumes, however, were only comparable for ESV (46.6 +/- 25.3 mL vs 48.8 +/- 37.0 mL, respectively; P =.13) but differed greatly for LVEF (35% +/- 13% vs 65% +/- 19%, respectively; P <.001), EDV (69.1 +/- 28.6 mL vs 125.9 +/- 53.4 mL, respectively; P <.001) and stroke volume (22.4 +/- 9.9 mL vs 77.1 +/- 33.7 respirations; P <.001). Moreover, Bland-Altman analysis showed the clinical noninterchangeability between these techniques for the measurement of hemodynamic parameters.
Conclusion:
Measurement of hemodynamic parameters with nonfluoroscopic mapping of the left ventricle is feasible and safe. The system provides data that strongly correlate but that are in clinical disagreement with angiographic data. Therefore the interchangeability of these techniques may be questioned.
More Related Videos
07:11Morphological and Functional Assessment of the Right Ventricle Using 3D Echocardiography
Published on: October 28, 2020
08:19Transthoracic Echocardiography to Assess Post-Resuscitation Left Ventricular Dysfunction After Acute Myocardial Infarction and Cardiac Arrest in Pigs
Published on: July 12, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...