Counting of single protein molecules at interfaces and application of this technique in early-stage diagnosis
1Institut für Analytische Chemie, Chemo- und Biosensorik, Universität Regensburg, Germany.
Abstract:
The fluorescence-based detection and counting of single protein molecules after specific binding to antibodies at interfaces is presented. A diode laser was used as the excitation source. The unspecific binding at the interface has been reduced to a level of only 0.1% of the maximum signal level. At present, the detection limit of this molecule-counting process is in the range of 10(-17) mol/L, and the dynamic range of the signal corresponds to 7 orders of magnitude of antigen concentration, but these values are not limiting. As a preliminary application in early-stage diagnosis, we have investigated the detection of a single cardiac actin molecule in human plasma, which is of interest in myocardial infarction diagnosis.