Measuring Spatially- and Directionally-varying Light Scattering from Biological Material
Published on: May 20, 2013
Universal opacity standard for Scheimpflug photography
1Research Laboratory, St. Erik's Eye Hospital, Karolinska Institutet, Stockholm, Sweden. wen@brain.ste.ki.se
Purpose:
To develop a universal standard for comparison of the results from different studies using Scheimpflug photography.
Material And Method:
A suspension including polystyrene spheres was selected as the universal standard. The polystyrene suspension was diluted into 10 concentrations by adding spectroscopic quality water. An artificial eye chamber was designed to hold the diluted suspension for photography. Each concentration was photographed by both a Topcon SL-45 slit-lamp camera and a Nidek EAS-1000 system. The intensity of back-scattered light was analyzed statistically.
Results:
The relationship between the concentration of the polystyrene suspension and the internal gray scale step within the Topcon SL-45 camera can be described by a second-order polynomial. The relationship between the concentration of the polystyrene suspension and the measurement of the Nidek EAS-1000 system can also be expressed by a second-order polynomial. The relationship between the Topcon SL-45 camera and the Nidek EAS-1000 system is linear.
Conclusions:
The results of lens nuclear opacity measured with different Scheimpflug cameras can be compared using the method established in the study. The linear relationship between the Topcon SL-45 and Nidek EAS-1000 systems can be used to convert measurements from one system to the other.
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...