Related Experiment Video
Updated: Oct 11, 2026

Design and Development of a Model to Study the Effect of Supplemental Oxygen on the Cystic Fibrosis Airway Microbiome
Published on: August 3, 2021
Estimate of gas transfer rates of an intravascular membrane oxygenator
T Kanamori1, M Niwa, H Kawakami
1Department of Polymer Engineering, National Institute of Materials and Chemical Research, Tsukuba, Ibaraki, Japan.
Abstract:
An asymmetric hollow fiber membrane was prepared from a newly synthesized fluorinated aromatic polyimide (6FDA-6FAP) by using a dry/wet phase inversion process. The membrane was used in a membrane oxygenator over a long period of time. In this study, the potential of the membrane for intravascular membrane oxygenation (IVOX) was studied in respect to oxygen transfer. The gas permeance of the membrane and three commercially available hollow fiber membranes for membrane oxygenators was measured in a gas-gas system and a gas-liquid system and discussed relative to the membrane structures. The oxygen transfer rates of the IVOX devices using these four membranes were estimated by a mathematical kinetic model, with the oxygen permeance measured in the gas-liquid system. The results showed that the device using the 6FDA-6FAP hollow fiber membrane has the highest oxygen transfer rate and is believed to be applicable to IVOX. The methods to determine oxygen transfer rate of a hollow fiber membrane and the mathematical kinetic model, are useful for developing a hollow fiber membrane and a device for oxygenation.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Bioreactor Controls-II
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Respiratory Volumes and Capacities
Maxwell-Boltzmann Distribution: Problem Solving
This distribution function f(v) is defined by saying that the expected number N (v1,v2) of particles with speeds between v1 and v2 is given by
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.