Related Experiment Videos
High-speed spectroscopic imaging for cancellous bone marrow R(2)* mapping and lipid quantification
L Hilaire1, F W Wehrli, H K Song
1University of Pennsylvania Medical Center, Philadelphia, PA 19104, USA.
Magnetic Resonance Imaging
|October 12, 2000
Summary
A new interleaved multiple-gradient-echo chemical shift imaging (IMGE-CSI) technique accurately quantifies lipids and measures R(2)* in bone marrow. This advanced method offers high resolution and reproducibility for in-vivo studies.
Area of Science:
- Magnetic Resonance Imaging
- Biomedical Engineering
- Quantitative Imaging
Background:
- Accurate in-vivo lipid quantification and R(2)* measurement are crucial for assessing bone marrow composition.
- Existing methods like Dixon-type techniques have limitations in spectral approximation.
- High-resolution imaging with fast scan times is essential for clinical applicability.
Purpose of the Study:
- To design, implement, and evaluate an interleaved multiple-gradient-echo chemical shift imaging (IMGE-CSI) technique.
- To achieve high-resolution lipid quantification and R(2)* measurement in cancellous bone marrow in-vivo.
- To compare IMGE-CSI performance against existing methods and assess its reproducibility.
Main Methods:
- Development and implementation of the IMGE-CSI technique at 1.5T and 4T.
- Utilizing spectroscopic bandwidth extension by interleaving gradient-echo trains.
- Employing multi-line spectral curve fitting for R(2)* determination.
Main Results:
- IMGE-CSI accurately measures true fat volume fraction, outperforming Dixon-type methods.
- High correlation (r(2) > 0.99) was observed between IMGE-CSI volume fractions and volumetry.
- The technique quantifies regional bone marrow composition with 2.5 x 2.5 x 5 mm(3) resolution.
- R(2)* measurements in cancellous bone marrow showed good agreement with time-domain methods and superiority in mixed constituents.
- Excellent inter-scan reproducibility (1% coefficient of variation) was demonstrated.
Conclusions:
- IMGE-CSI enables highly accurate in-vivo lipid quantification.
- The technique provides precise measurement of cancellous bone marrow R(2)*.
- IMGE-CSI achieves resolutions and scan times compatible with clinical protocols.