Objective:
To examine whether the interface pressure (IP) relief provided by alternating pressure air mattresses (APAMs) is matched with maintenance of tissue perfusion over the points of contact by measuring transcutaneous oxygen and carbon dioxide (tcPO2, tcPCO2).
Design:
Comparative analysis of 2 APAMs with a 2-parameter continuous time-based method for quantifying pressure relief (PR) and transcutaneous gas measurement for assessing tissue perfusion.
Setting:
Rehabilitation research facility in a university hospital.
Participants:
Eleven able-bodied adult postgraduate student volunteers.
Main Outcome Measures:
Two full-replacement APAM systems were used. For each mattress the mean maximum and minimum interface pressures; mean peak air pressures in the mattresses; interface pressure durations below 30, 20, and 10 mmHg over a 60-minute period; mean maximum tcPCO2 and minimum tcPO2; and mean area under the tcPO2 and tcPCO2 curves were measured for each subject.
Results:
IP on the sacrum was held below thresholds of 30, 20, and 10 mmHg longer on a 2-cell, low pressure system than on a 3-cell, high pressure system (p < .001). Integrated over time, tcPO2 levels also indicated that the 2-cell system retained oxygen levels closer to the unloaded baseline than did the 3-cell system (p < .01). tcPCO2 levels did not rise significantly (p > 0.1) compared with the baseline measurement in both mattresses.
Conclusions:
PR was sensitive to the design of the APAM, especially its inflation pressure, cycle time, and inflation sequence. If future trials demonstrate that PR values and transcutaneous blood gas measurements correlate significantly with the clinical incidence of pressure sore formation, then this technique may prove useful in assessing the effectiveness of alternating pressure support surfaces.
More Related Videos
09:31Normothermic Negative Pressure Ventilation Ex Situ Lung Perfusion: Evaluation of Lung Function and Metabolism
Published on: February 14, 2022
06:45Development of a Benchtop Model for Evaluating the Compatibility of Wound Dressing Materials with Negative Pressure Wound Therapy Systems
Published on: May 2, 2025
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Pressure Variation in a Fluid at Rest
When measuring pressure at two different levels within the fluid, the difference in pressure...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...