A validation of the Mobil O Graph (version 12) ambulatory blood pressure monitor
C R Jones1, K Taylor, P Chowienczyk
1Fetal Health Research Group and Department of Obstetrics and Gynaecology, Centre for Cardiovascular Research and Medicine, Guy's, King's and St Thomas' School of Medicine, St Thomas' Hospital, London SE1 7EH, UK.
Objective:
To assess the clinical accuracy of the Mobil O Graph (version 12) ambulatory blood pressure monitor in an adult population.
Methods:
The accuracy of the device was assessed by predefined criteria (British Hypertension Society, BHS) in 85 subjects recruited from the patients and staff in a teaching hospital. A series of same-arm sequential blood pressure measurements were taken: first two observers taking simultaneous mercury readings, followed by a reading with the Mobil O Graph ambulatory monitor. A total of seven readings were taken from each subject in the sitting position. The data were then analysed according to the BHS protocol and the criteria of the Association for the Advancement of Medical Instrumentation (AAMI).
Results:
The Mobil O Graph ambulatory monitor fulfilled the criteria of the BHS protocol, achieving a grade B for systolic blood pressure (SBP) and a grade A for diastolic blood pressure (DBP). The mean differences were -2+/-8 mmHg for SBP and -2+/-7 mmHg for DBP. The device therefore also passed the AAMI standard (the mean to be within 5+/-8 mmHg).
Conclusion:
The Mobil O Graph ambulatory monitor performed in a satisfactory manner according to the BHS and the AAMI criteria and can therefore be recommended for clinical use in the general population.
More Related Videos
14:28Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure