Related Experiment Video
Updated: Oct 11, 2026

Veno-Venous Extracorporeal Membrane Oxygenation in a Mouse
Published on: October 24, 2018
Quantification of recirculation by thermodilution during venovenous extracorporeal membrane oxygenation
C Sreenan1, H Osiovich, P Y Cheung
1Department of Pediatrics, University of Alberta, Edmonton, Canada.
Purpose:
The aim of this study was to determine whether recirculation could be quantified by a thermodilution technique during venovenous (VV) extracorporeal membrane oxygenation (ECMO) in a rabbit model.
Methods:
Five New Zealand white rabbits, mean weight, 4.5 (range, 3.7 to 5.7) kg, were anesthetized, instrumented, cannulated with a double-lumen catheter, and placed on VV ECMO. Serial injections of ice-cold saline were performed at the arterial arm of the circuit, and the resultant temperature change at various pump flows was measured at the venous arm of the circuit using a thermistor-tipped catheter and a cardiac output computer. Results were compared with the respective 100% recirculation measured with all the circuit flow passing through the bridge.
Results:
Using linear regression, recirculation percentage could be calculated as: 19 + 0.1 x pump flow (R2 = 0.81, P < .005). Recirculation correlated positively with pump flow. Variability between results at each flow was less than 10%.
Conclusions:
Recirculation can be quantified during VV ECMO by measuring the change in temperature in the venous arm using a cardiac output computer after injection of a known quantity of ice-cold saline in the arterial side of the circuit. The effect of interventions to reduce recirculation can be assessed conveniently and reliably.
Related Concept Videos
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system as it...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.