Technique factors and image quality as functions of patient weight at abdominal CT
W Huda1, E M Scalzetti, G Levin
1Department of Radiology, SUNY Upstate Medical University, 750 E Adams St, Syracuse, NY 13210, USA. hudaw@mail.upstate.edu
Purpose:
To investigate how changes in kilovolt peak and milliampere second settings, and patient weight affect transmitted x-ray energy fluence and the image contrast-to-noise ratio (CNR) at abdominal computed tomography (CT).
Materials And Methods:
Cylinders of water were used as patient models, and x-ray spectra, including x-ray tube potentials of 80-140 kVp, were investigated. The mean photon energy and energy fluence transmitted through water cylinders with varying diameters and the image contrast for fat, muscle, bone, and iodine relative to water were determined. The effect of changing the x-ray tube potential on CNR also was investigated.
Results:
At a constant kVp, increasing patient weight from 10 kg to 120 kg reduced the transmitted energy fluence by two orders of magnitude. Changing the x-ray tube potential from 80 kVp to 140 kVp increased the mean photon energy from approximately 52 keV to approximately 72 keV and thus reduced the image contrast relative to water by 12% for muscle, 21% for fat, 39% for bone, and 50% for iodine (approximate reduction values). Increasing the x-ray tube potential from 80 kVp to 140 kVp increased the CNR by a factor of 2.6 for muscle and by a factor of 1. 4 for iodine.
Conclusion:
With changes in patient weight at abdominal CT, x-ray tube potentials must be varied to maintain a constant detector energy fluence. Increasing the x-ray tube potential generally improves CNR.
More Related Videos
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
06:50Troubleshooting FoCUS Image Acquisition: Patient Positioning, Transducer Manipulation, and Image Optimization
Published on: March 3, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography