Device eccentricity: postmagnet behavior of DDDR pacemakers with automatic threshold tracking
1St. Jude Medical, CRMD, Sylmar, California 91342, USA. plevine@pacesetter.com
Pacing and Clinical Electrophysiology : PACE
|November 4, 2000
Abstract:
The normal pacing system function associated with the AutoCapture Pacing System algorithm triggered concern on the part of the clinical staff caring for the patient when the initiation of a threshold search sequence was detected during a transtelephonic follow-up evaluation. The analysis of the rhythm demonstrates that the behavior of the system is normal and consistent with the design of the AutoCapture algorithm. As a variety of new algorithms are introduced, similar unexpected but normal behaviors can be anticipated.