Introduction:
The safety and efficacy of medications stored on air medical helicopters may be adversely affected by extreme temperatures. The purpose of this study was to determine whether temperatures inside an air medical helicopter drug box were within the U.S. Pharmacopeia recommendations for controlled room temperature. This is defined as a temperature between 15 degrees and 30 degrees C (59 degrees and 86 degrees F) with a mean kinetic temperature of less than 25 degrees C (77 degrees F). An additional goal was to determine whether time/temperature indicator labels can reliably monitor mean kinetic temperatures.
Methods:
Temperatures were monitored with miniature electronic temperature recorders and color-changing time/temperature indicator labels.
Results:
The mean kinetic temperatures for the summer and winter periods were 25.1 degrees C (77.2 degrees F) and 12.7 degrees C (54.8 degrees F), respectively. In the summer, the electronic recorders logged temperatures exceeding 25 degrees C (59 degrees F) 37% of the time and more than 30 degrees C (86 degrees F) 6% of the time. In the winter, temperatures less than 15 degrees C (59 degrees F) were recorded 83% of the time. The mean kinetic temperatures obtained from the electronic recorder and the time/temperature indicator labels differed by less than 0.7 degree C (1.3 degrees F). The results show that medications on an air medical helicopter are subject to temperatures out of the recommended range and that time/temperature indicator labels can reliably monitor mean kinetic temperatures.
More Related Videos
06:42Optimization of Breast Biopsy and Mastectomy Sample Collection Procedures for Biobanking, Personalized Medicine, and Research Applications
Published on: September 2, 2025
08:15Innovative Strategies for Organ Preservation in Heart Transplantation: Uniform Cooling Preservation and Ex-situ Normothermic Perfusion
Published on: November 28, 2025
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Sample Handling
Samples should be transported carefully from collection points to the laboratory. They should be properly sealed and clearly labeled to prevent cross-contamination. To preserve the sample integrity, optimal temperature conditions during transport are essential. This could involve using...
Inhaled Medications
Pharmaceutical Alternatives: Stability-Related Therapeutic Nonequivalence