Mechanistic investigation of ionization suppression in electrospray ionization
Related Experiment Video
Updated: Oct 10, 2026

Sample Preparation for Probe Electrospray Ionization Mass Spectrometry
Published on: February 19, 2020
R King1, R Bonfiglio, C Fernandez-Metzler
1Department of Drug Metabolism, West Point, Pennsylvania 19468, USA. richard_king@merck.com
Abstract:
We show results from experiments designed to determine the relative importance of gas phase processes and solution phase processes into ionization suppression observed in biological sample extracts. The data indicate that gas phase reactions leading to the loss of net charge on the analyte is not likely to be the most important process involved in ionization suppression. The results point to changes in the droplet solution properties caused by the presence of nonvolatile solutes as the main cause of ionization suppression in electrospray ionization of biological extracts.